Quantitative Determination of Meloxicam in bulk and in tablet by UVSpectrophotometry
Keywords:
Meloxicam, UV spectrophotometry, validation, area under curve, first order derivationAbstract
Three simple, precise and economical UV-Spectrophotometric methods have been established for the quantification of
meloxicam in bulk and in tablet. Zero order spectrophotometer determination was done at 269 nm in method I; in method II, Area under curve was done in the wavelength range of 254-279 nm; whereas, zero order spectrum was derivatized into first order (??=2) at wavelength 275 nm in method III. In method I and II, meloxicam obeyed linearity in the concentration range of 5-30 g/ml while in method III, 50-300 ?g/ml, respectively with r2 obtained as 0.999, 0.998 and 0.999. Calibration curves were plotted using instrument response between particular wavelengths and concentrations of analyte in the solution. The proposed methods were successfully applied for the determination of meloxicam in commercial tablets and amount found to be 99.81%, 100.15% and 100.07% respectively. The proposed methods were validated as per ICH guidelines for accuracy, precision, repeatability and ruggedness.
Downloads
References
O’ Neil MJ. The Merck Index: An Encyclopedia of
chemicals, drugs and biologicals, 14th Ed. New Jersey:
Published by Merck Research Laboratories, Division of
Merck and Co., Inc. Whitehouse station.1996; 12.
Rigato H, Mendes G, Borges N, et al. Meloxicam
determination in human plasma by HPLC coupled with
tandem mass spectrometry (LC-MS-MS) in Brazilian
bioequivalence studies. Inter. J. Clinc. Pharm. Thera.
; 44(10): 489-498.
El-Ries A, El-Dars F, Attia F, et al. Potentiometric and
HPLC methods for the determination of meloxicam in
bulk and drug formulations. Insight Pharm. Sci. 2012,
(1).
Arayne M, Sultana N, Siddiqui F. A new RP-HPLC
method for analysis of meloxicam in tablets. Paki. J.
Pharm. Sci. 2005; 18(1): 58-62.
Desai N, Amin. Stability indicating HPTLC
determination of meloxicam. Ind. J. of Pharm. Sci.
; 70(5): 644-647.
Starek M, Krzek J. TLC Determination of meloxicam in
tablets and after acidic and alkaline hydrolysis. Acta.
Pol. Pharm. 2012; 69(2): 225-235.
Master S, Mehta R, Bhatt K, et al. First order derivative
spectroscopic method for simultaneous estimation of
meloxicam and paracetamol in their combined dosage
form. Inter. J. of Res. in Pharm. and Chem. 2011; 1(2):
-198.
Nemutlu E, Kir S. Validated determination of meloxicam
in tablets by using UV Spectrophotometry. J. of Faculty
of Pharm., 2004; 24(1): 13-24.
Sawant R, Joshi R, Kawade D, et al. Development and
validation of spectrophotometric methods for
simultaneous estimation of paracetamol and meloxicam
in pure and tablet dosage form Der Pharmacia Letters.
; 2(2): 471-478.
Khan F, Lohiya R, Umekar M, Development of UV
Spectrophotometric method for the simultaneous
estimation of meloxicam and paracetamol in tablet by
simultaneous equation, absorbance ratio and absorbance
correction method. Inter. J. of Chem Tech Res. 2010;
(3): 1586-1591.
Text on Validation of analytical procedures, International
Conference on harmonization. ICH guidelines Q2A;
Validation of analytical Procedures; Methodology.
International Conference on harmonization, ICH
guidelines Q2B;1996.
Published
How to Cite
Issue
Section

This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
Copyright © Author(s) retain the copyright of this article.